Ledford's state-of-the-art
facility was designed with dimensional calibration in mind.
Opened in 1998, the laboratory has double wall construction with
a four inch air envelope allowing optimum environmental control.
Temperature and humidity are controlled by a CSI computer system
which samples every thirty seconds and adjusts every three minutes
as needed. A 25 inch slab floor, R-60 insulation and a .0101 micron
filter system ensure the conditions necessary for measurements
in the millionths.
The lab is equipped with state-of-the-art computerized technical
precision equipment including coordinate measuring machines, a
laser interferometer system and micro-process or digital height
gages. This equipment speeds service and assures precise measurement
accuracy.
|